- ICP-OES Inductively Coupled Plasma Optical Emission Spectrometer ,ICP-5000
ICP-5000
Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES), is a well- established analytical technique that used to detect the presence of trace metals in an analyte. The analyte sample is introduced into the ICP as an aerosol that is carried into the center of the plasma – ionized argon gas. The superheated argon gas excites ions and atoms of the element in the sample that emits detectable amounts of light at characteristic wavelengths. The intensity of the emission is propor tional to the concentration of the element detected. The ICP-5000 can be programmed for both radial and axial view. The less sensitive radial view was preferred for the higher concentration elements such as aluminum, calcium, iron, magnesium, potassium and sodium in environmental samples while the more sensitive axial view was preferred for the lower concentration elements such as the transition series. The most effective systems of ICP-5000 allow the plasma to work in dual view in a single analysis, providing the highest upper linear ranges and the best detection limits.
Includes a 12-roller 3-channel, computer-controlled peristaltic pump with automatically adjustable speed 0-125rpm. The ICP-5000 is supplied with a standard glass concentric nebulizer and a glass cyclonic spray chamber, while HF-resistant nebulizer and spray chamber are supplied for high HF/salt content analytes.
pngICP SystemThe ICP system comprises of a free-running solid- state RF generator, a load coil, a demountable quartz torch and automatically controlled gas flow. The RF generator is continuously adjustable from 750 to 1350 watts (Dual), maximum up to 1600 watts (Radial). The coupling efficiency is greater than 80% with ? 0.1% variation in output power stability and ?0.01% variation in output frequency stability. The RF generator produces an oscillating electromagnetic field at a frequency of 27.12MHz. This radiation is directed to the load coil through which delivered to the torch. The argon gas flowing through the torch will form a plasma in the RF field. Plasma argon, auxiliary argon and nebulizer argon are regulated in a fully integrated gas flow and automatically controlled by MFC in 0.01ml/min increments.
pngSpectrometerPolychromator : 2D Echelle Grating, 52.91 lines per mm and a blaze angle of 63.4 degrees. Thermostatted optics: The entire optical system is encapsulated in a thermostatted enclosure 36?±0.5? and purged with argon or nitrogen gas (UV region). Recalibration: Spectra of Carbon, Nitrogen and argon elements are used for automatic wavelength calibration during ever y ignition of the plasma. Do not require either calibration solution or mercur y/neon lamp that is free from calibration consumables or preheating of the lamp.
pngDetectorThe ICP-5000 features a megapixel Cryogenic area-array Charge-coupled Device (CCD) detector with unique back-illuminated technique results in average quantum efficiency greater than 75% in UV region. Three-level Thermo Electric Coolers (TEC) are equipped that can drive the cryogenic temperature down to -45?, cooling time<3min. Anti-blooming of the detector is functioned in pixel level.